MF11 – 3.08 Installation, Operation & Maintenance Instructions 1200°C Thermocouple Calibration Furnace type PTC 12/20/150 This manual
PTC 10 MF11 – 3.08 4.11 Stopping the Timer To stop the timer at any time while it is running, change the StAt parameter to OFF. This is the sam
PTC MF11 – 3.08 11 5.0 MAINTENANCE 5.1 General Maintenance No routine maintenance is required other than the occasional replacement of consuma
PTC 12 MF11 – 3.08 6.0 REPAIRS & REPLACEMENTS 6.1 Safety Warning – Disconnection from Supply Always ensure that the furnace is disconnecte
PTC MF11 – 3.08 13 6.5 Thermocouple Replacement Disconnect the furnace from the supply, and remove the furnace back panel (opposite the working
PTC 14 MF11 – 3.08 7.0 FAULT ANALYSIS A. Furnace Does Not Heat Up 1. The HEAT light is ON The heating element has failed Check also th
PTC MF11 – 3.08 15 8.0 CIRCUIT DIAGRAM 8.1 Circuit Diagram 9.0 FUSES & POWER CONTROL 9.1 Fuses Supply Fuses: 32mm x 6mm t
Carbolite, Parsons Lane, Hope, Hope Valley, S33 6RB, England. Telephone: +44 (0)1433 620011 Fax: +44 (0)1433 621198 E-mail: [email protected] The
PTC 2 MF11 – 3.08 1.0 SYMBOLS & WARNINGS 1.1 Switches and Lights 1.2 Warning Symbols Instrument switch: when the instrum
PTC MF11 – 3.08 3 2.0 INSTALLATION 2.1 Unpacking Unpack the furnace and remove the transit clamp from the working end of the furnace as follow
PTC 4 MF11 – 3.08 3.0 OPERATING INSTRUCTIONS 3.1 Operating Cycle The furnace is fitted with an instrument switch. The switch cuts off powe
PTC MF11 – 3.08 5 diameter is 7.5mm - larger diameters may experience too much heat conduction and give an incorrect calibration. Insert the the
PTC 6 MF11 – 3.08 4.0 CONTROLLER OPERATION 4.1 Eurotherm 2132 When switched on, the controller lights up, goes through a short test routine,
PTC MF11 – 3.08 7 4.5 Altering the Ramp Rate It is possible to limit the rate of heating by setting a ramp rate only if the timer feature is no
PTC 8 MF11 – 3.08 4.7 Operation With the Timer The 2132 can be used as a process timer allowing timed heating or timed delay, according to the
PTC MF11 – 3.08 9 4.8 Setting the Timer Mode Scroll to tm.OP; use or to view and alter the mode. The mode shows as OPt.1 to OPt.5. It i