MF10 - 3.08 Installation, Operation & Maintenance Instructions Combustion Tube Furnaces CFM models This manual is for the guidance of
CFM 10 MF10 - 3.08 5.5 Thermocouple Replacement Disconnect the furnace from the supply, and remove the end panel at the thermocouple end. Mak
CFM MF10 - 3.08 11 5.7 Element Fitting and Replacement - 1400°C models Replacements: read the section on element aging (0). If at any time the
CFM 12 MF10 - 3.08 6.0 FAULT ANALYSIS A. Furnace Does Not Heat Up 1. The HEAT light is ON The heating element has failed Check also th
CFM MF10 - 3.08 13 7.0 CIRCUIT DIAGRAMS Safety switches: where the furnace rating is over 16A, there is a single switch in the contactor coil cir
CFM 14 MF10 - 3.08 7.3 1400°C Model with Auxiliary Heater (earlier models) 7.4 Higher Voltages For 254V (or above) an isola
CFM MF10 - 3.08 15 FUSES & POWER SETTINGS 7.5 Fuses F1-F3: Refer to the circuit diagrams. F1 Internal supply fuses Fitted if supply cab
Carbolite, Parsons Lane, Hope, Hope Valley, S33 6RB, England. Telephone: +44 (0)1433 620011 Fax: +44 (0)1433 621198 E-mail: [email protected] The p
CFM 2 MF10 - 3.08 1.0 SYMBOLS & WARNINGS 1.1 Switches and Lights 1.2 Warning Symbols Instrument switch: when the instr
CFM MF10 - 3.08 3 2.0 INSTALLATION 2.1 Unpacking & Handling When unpacking or moving the furnace always lift it by its base. Do not lift i
CFM 4 MF10 - 3.08 2.5 Electrical Connections Connection by a qualified electrician is recommended. The furnaces covered by this manual requir
CFM MF10 - 3.08 5 3.0 OPERATION The instructions for operating the temperature controller are given in a separate manual. If the furnace is
CFM 6 MF10 - 3.08 3.3 General Operating Advice Heating element life is shortened by use at temperatures close to maximum. Do not leave the furn
CFM MF10 - 3.08 7 4.0 MAINTENANCE 4.1 General Maintenance No routine maintenance is required other than the occasional replacement of consumab
CFM 8 MF10 - 3.08 All models. Occasionally the power limit is set to zero to permit demonstration of the controls without the heating elements t
CFM MF10 - 3.08 9 5.0 REPAIRS & REPLACEMENTS 5.1 Safety Warning – Disconnection from Supply Always ensure that the furnace is disconnected