MF03 – 3.13 Installation, Operation & Maintenance Instructions 1000°C - 1200°C Tube furnaces types MTF, CTF This manual is for the guid
MTF, CTF 10 MF03 3.13 Check the cylindrical case and clean out as appropriate. Remove any insulation sleeving from the tails of the old elemen
MTF, CTF MF03 3.13 11 6.0 FAULT ANALYSIS A. Furnace Does Not Heat Up 1. The HEAT light is ON The heating element has failed Check also
MTF, CTF 12 MF03 3.13 7.0 CIRCUIT DIAGRAMS Safety Switches type A: a 2-pole Heater Switch is fitted directly in the element circuit in models u
MTF, CTF MF03 3.13 13 7.4 3-phase without neutral (delta) 7.5 3-phase without neutral (star – e.g. 380 to 415V) 7.6 Higher Vo
MTF, CTF 14 MF03 3.13
MTF, CTF MF03 3.13 15 8.0 FUSES & POWER SETTINGS 8.1 Fuses F1-F3: Refer to the circuit diagrams. F1 Internal supply fuses Fitted if s
Carbolite, Parsons Lane, Hope, Hope Valley, S33 6RB, England. Telephone: +44 (0)1433 620011 Fax: +44 (0)1433 621198 E-mail: [email protected] The p
MTF, CTF 2 MF03 3.13 1.0 INTRODUCTION 1.1 Switches and Lights 1.2 Warning Symbols Instrument switch: when the instrument sw
MTF, CTF MF03 3.13 3 2.0 INSTALLATION 2.1 Unpacking & Handling When unpacking or moving the furnace always lift it by its base or by both
MTF, CTF 4 MF03 3.13 2.4 Electrical Connections Connection by a qualified electrician is recommended. All furnaces covered by this manual may
MTF, CTF MF03 3.13 5 3.0 OPERATION The instructions for operating the temperature controller are given in a separate manual. If the furnac
MTF, CTF 6 MF03 3.13 3.4 Tube Life A ceramic work tube may be cracked if workpieces are inserted too quickly or at temperatures below 900°C (wh
MTF, CTF MF03 3.13 7 4.0 MAINTENANCE 4.1 General Maintenance No routine maintenance is required. The outer surfaces may be cleaned with a damp
MTF, CTF 8 MF03 3.13 5.0 REPAIRS & REPLACEMENTS 5.1 Safety Warning – Disconnection from Supply Always ensure that the furnace is disconnec
MTF, CTF MF03 3.13 9 5.5 Thermocouple Replacement The coverings and guards which must be removed to gain access to the thermocouple depend on t