MF22 – 3.09 Installation, Operation & Maintenance Instructions 1700-1800°C Chamber Furnaces types RHF & HTF This manual is for th
RHF & HTF 10 MF22 3.09 5.6 Element Installation & Replacement Safety Warning – Molybdenum Disilicide elements. The elements form a glaze
RHF & HTF MF22 3.09 11Use of Clip Tool larger clips require clip tool, comprising two levers smaller clips do no require clip tool Elemen
RHF & HTF 12 MF22 3.09 6.0 FAULT ANALYSIS A. Furnace Does Not Heat Up 1. The HEAT light is ON An ohm meter applied to the element c
RHF & HTF MF22 3.09 137.0 CIRCUIT DIAGRAM 7.1 Single Phase Example - Manual Door & Electric Door Versions
RHF & HTF 14 MF22 3.09 8.0 FUSES & POWER CONTROL 8.1 Fuses F1-F2: Refer to the circuit diagrams. F1 Internal supply fuses Fitted if
RHF & HTF MF22 3.09 159.0 SPECIFICATIONS Carbolite reserves the right to change specifications without notice. 9.1 Models Covered by this M
RHF & HTF 16 MF22 3.09
RHF & HTF Carbolite, Parsons Lane, Hope, Hope Valley, S33 6RB, England. Telephone: +44 (0)1433 620011 E-mail: [email protected] The products cov
RHF & HTF 2 MF22 3.09 1.0 SYMBOLS & WARNINGS 1.1 Switches and Lights 1.2 Warning Symbols Instrument switch: when the i
RHF & HTF MF22 3.09 32.0 INSTALLATION 2.1 Unpacking & Handling When unpacking or moving the furnace always lift it by its base. Never l
RHF & HTF 4 MF22 3.09 2.4 Electrical Connections Connection by a qualified electrician is recommended. The smaller models are designed on
RHF & HTF MF22 3.09 53.0 OPERATION The instructions for operating the temperature controller are given in a separate manual. If the furna
RHF & HTF 6 MF22 3.09 to the progressive shrinkage of the insulation materials. Cracks are not usually detrimental to the functioning or the s
RHF & HTF MF22 3.09 73.6 Thermocouples - Warning (1) The output from 1700-1800°C thermocouples when used regularly at temperatures greater th
RHF & HTF 8 MF22 3.09 4.0 MAINTENANCE 4.1 Routine Maintenance No routine maintenance is required except for the replacement of consumable
RHF & HTF MF22 3.09 95.0 REPAIRS & REPLACEMENTS 5.1 Safety Warning – Disconnection from Supply Always ensure that the furnace is discon